Ask a better questionLesson 04 of 30 · 5 min
What does your timer miss?
Explain how resolution and calibration affect a measured difference.
A measurement procedure turns an event into a number. Its resolution is the smallest change it can distinguish. A timer rounded to the nearest 10 ms cannot reliably separate two very short durations that fall into the same rounding bin.
Calibration compares an instrument with a reference. An offset that adds 5 ms to every reading makes absolute times wrong; it may cancel in a difference if it is truly the same in both conditions. A changing offset will not cancel so neatly.
The example starts from invented reference durations and rounds them. Repeating a rounded reading cannot recover information the instrument never recorded. Timing a batch can help with resolution, but dividing batch time by operations estimates an average under that batching procedure, not each operation's latency.
Can two distinct reference durations produce the same displayed reading?
Write a prediction (optional)
Run the prepared example
This page puts the example in the Python editor. Select Run Python with the code unchanged.
Authored instrument model
A saved draft is open. To run the published example, open Editor tools and select Restore example. Recover draft brings your code back.
Execution error
Technical error details
What to compare
Find 11, 12, 13 and 14 in Reference (ms) and compare their entries in Displayed (ms). Then compare the gap between the first two reference times with the gap between their offset readings.
Your result
Result tools
Four different reference durations round to 10 ms. A constant offset changes absolute times by 5 ms and leaves the gap between any two unrounded times unchanged.
Published example output
The site ran the original code to produce this reference. Build environment ↗
Reference (ms): [11 12 13 14 21 22] Displayed (ms): [10. 10. 10. 10. 20. 20.] With offset (ms): [16 17 18 19 26 27]
Four different reference durations round to 10 ms. A constant offset changes absolute times by 5 ms and leaves the gap between any two unrounded times unchanged.
Change and run again
In the Python editor, change resolution_ms from 10 to 1 and select Run Python. Compare Displayed (ms) with the first run. Next change offset_ms = 5 to offset_ms = np.array([0, 1, 2, 3, 4, 5]) and run again. Compare the gap between the first two entries in Reference (ms) and With offset (ms). Does a varying offset preserve that gap? Return to the Python editor
What this example supports. Rounding is an authored instrument model. Real timers can also have overhead, scheduling delays and drift.
Try a fresh case
A sensor consistently reads 2 °C too high. Repeating it 1,000 times will mainly do what?
Hints for this question
- Separate a random fluctuation from a systematic offset.
- Ask what changes between repetitions.